Struers Accutom-5 Mode D'emploi page 48

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Error Messages
Display Message
X-MIN sensor not activated
X-MIN sensor not deactivated
Encoder error
Cutting motor overloaded
Machine Problems
The recirculation pump is not
working correctly
44
Accutom-5
Instruction Manual
Explanation
The inductive sensor cannot detect
the end position of the X-table
The inductive sensor for the X-posi-
tion cannot be deactivated
The Y-table does not move, or the
encoder is defective
The main motor has been exposed
to a high load for a long time
Explanation
The water level in the recirculation
tank is too low
The water outlet is clogged
The recirculation pump is
contaminated with cutting residue
Action
Check for mechanical blockage of
the table. If that is not the case, call a
Struers service technician
Check for mechanical blockage of
the table. If that is not the case, call a
Struers service technician
Check for mechanical blockage of
the table. If that is not the case, call a
Struers service technician
Wait until the motor has cooled down
"Cutting motor ready after overload"
will be shown on the display. Adjust
process parameters to avoid
repetitions
Action
Check that there is sufficient water in
the recirculation tank
Remove the cooling tubes and run
the recirculation pump.
Flush the cooling tubes with clean
water before replacing
Flush the pump with clean water by
gently pressing fresh water into the
inlet tube