Error Messages
Display Message
Searching for XY-reference position
Reference pos. not found, close
cover
Cover open!
Process stopped by open cover
Process running
Y-stop position outside cutting range The set stop position is outside the
Y-stop position less than Y-relative
Feed less than 90% of pre-set value
Cutting stopped, feed too low
Cutting method not selected
Dressing range exceeded
Accutom-5
Instruction Manual
4. Trouble-Shooting
Explanation
The machine searches for the initial
reference positions
Accutom-5 was started with the
cover open and did not search for
the reference position
You are trying to start the cutting
process or a dressing while the cover
is open
You have opened the cover during a
cutting process or a dressing
process
You are trying to start two different
processes at the same time
maximum range in the Y-direction.
(max. 105.0 mm)
You try to start a cutting process
when the relative Y-position is larger
than or equal to the Y-stop
The force has exceeded the force
limit (LOW, MEDIUM, HIGH) and the
feed speed is automatically reduced
The feed speed has been reduced to
less than 10% due to the force
exceeding the force limit
was pressed before a
START
cutting method was selected
The parameter Y-distance is set to a
value larger than the possible
movement of the table
Action
Wait for the search process to finish
Close the cover, wait for the
reference search to finish and press
START
again
Close the cover and press START
again
Close the cover and restart the
interrupted process
Wait for the previous process to
finish and try again
Adjust Y-stop position.
Position the specimen holder
correctly and/or change the Y-stop
position.
Decrease the feed speed or increase
the force limit next time you cut the
same material
Decrease the feed speed, increase
the force limit or change the cut-off
wheel and re-start the cutting
process
Select a method and press
again
START
Adjust the Y-distance and press
START again
43