Operation - Additional Instructions for LCB2500/2 only
Delete all data [NB] –
It is not possible to delete individual results
1. Move the rotary selector switch to the RCL position.
The code rcl is displayed.
2. Press the UP and DOWN keys together. The code
dEL is displayed.
3. Confirm that the data is no longer required by
pressing the SAVE key or abort by pressing any
other key. The code rcl is displayed.
Note: Once deleted, test data cannot be recovered
Print Results
1. Connect printer and the instrument with a serial
printer lead.
2. Move the rotary selector switch to the RCL position.
The code rcl is displayed.
3. Commence the printout by pressing the TEST key.
'Prn' is displayed Abort at any time by pressing and
holding the ENTER key. The code rcl is displayed.
Printer Set-up Mode
The instrument cannot respond to a busy signal given
by a printer, and therefore waits at the end of each line.
This wait time and the printer report language can be
changed.
1. Press and hold the backlight key then turn the
rotary selector switch from the OFF position to the
RCL position. The code Pdt is displayed.
2. Release the backlight key.
30
To change the Printer wait time
1. Toggle the TEST TYPE key to scroll through and
display the code Pdt.
2. Press the ENTER key. The current setting is
displayed.
3. Toggle the UP and DOWN keys until the required
setting is displayed.
4. To save the new setting, press the SAVE key. The
bleeper sounds and Std is displayed. To abort the
new setting, press the EXIT key.
To select the printer language
1. Toggle the TEST TYPE key to scroll though and
display the code lng.
2. Press the ENTER key. The current printer report
language is displayed as 1 (English) or 2 (as given
on the type label on the User Guide cover). AVO
Download Manager supplied with your instrument
enables the second printer language to be changed.
Refer to the instructions supplied with the disk.
3. Toggle the TEST TYPE key until the required
language setting is displayed.
4. To save the new setting, press the SAVE key. The
bleeper sounds and Std is displayed. To abort the
new setting, press the EXIT key.