The instrument displays an error code if it detects a problem during
startup or operation. See TABLE 5-1 for a brief description of the error
and proper corrective action.
Table 5-1. Troubleshooting Guidelines
PROBLEM
DESCRIPTION
DISPLAY
TOGGLES
BETWEEN:
TMP/ERR
Temperature out of range
AD/ERR
Sensor feedback error
EE/ERR
EEPROM inoperative
PRG/ERR
Program Memory
inoperative
RAM/ERR
RAM inoperative
TMR/ERR
Time or Clock Error
BTN/ERR
Button error (stuck down)
PWR/ERR
Unexpected Power Loss
Errors
LED/ERR
LED inoperative
VIB/ERR
Vibrator inoperative
UNK/ERR
Unknown Error
UND/ERR
Sensor Under range
Low battery warning
indicator (no alarm)
Low battery Alarm
indicator (horn sounds,
/ERR
LEDs activate
SNS/ERR
Sensor Error
ACTION
Contact MSA during warranty period
Contact MSA during warranty period
Contact MSA during warranty period
Contact MSA during warranty period
Contact MSA during warranty period
Contact MSA during warranty period
Contact MSA during warranty period
Contact MSA during warranty period
Verify operation of two alarm
LEDs in alarm
Verify operation of vibrator in alarm
Contact MSA during warranty period
Calibrate instrument
Remove from service
Remove from service
Contact MSA during warranty period.