Press the TEST button again:
The "√" that appears on the display:
indicates that the instrument passed the bump test.
remains for 24 hours, indicating that a bump test
was performed on this instrument within the last 24
hours.
NOTE: See FIGURE 3-1 for details
If If the "√" does not appear and and "ERR" is indicated:
• Check that the:
sensor inlet is not clogged
correct calibration cylinder is used to perform the bump
test
gas cylinder has not expired and is not empty
gas was applied at the appropriate time
gas tubing is seated in the front instrument case during
testing.
• Repeat the bump test process as needed.
If the "√" does not appear,calibrate the instrument and repeat the
Bump Test.