Setup and initial use
Check on delivery

4.2 Transport

CAUTION
Moving the
unpacked
instrument

4.3 Setting up the instrument base

Where?
82
Unpack the rotary evaporator and ensure that it is complete and
undamaged.
In case of damage, please inform Heidolph Instruments service
(see chapter 8.4, page 125).
Vibration and mechanical shocks!
Damage to the housing and mechanism of the instrument.
Avoid vibration and mechanical shocks.
Ensure that the heating bath is empty.
Remove all hose and cable connections to other instruments, the cooling
water/vacuum connection, and unplug the instrument from the AC power
socket.
Hold the underside of the instrument base.
Chemicals may be released into the atmosphere via the pump when handing
hazardous materials and distilling with solvents.
Ensure that an exhauster is available for the exhaust air of the pump.
Do not set up the rotary evaporator in a corrosive atmosphere.
Choose a level, firm base.
Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control
1.02