Auto Start Stop Function - Heidolph Laborota 4000 efficient Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

6.8 Auto Start Stop function

Stopping the Auto
Start Stop function
1.02
An acoustic signal indicates that the data record, including the pressure ramp
values, has been saved.
The programmed pressure ramps can be edited at any time by changing the
relevant parameters (see chapter 6.6.1, page 117, and see chapter 6.7.1,
page 117).
Use the Auto Start Stop function to automatically start and stop distillation.
Instrument is operational.
Evaporating flask immersion depth has been set (see chapter 4.5.7,
page 94).
Desired settings have been made.
1. Press Auto Start Stop (Fig. 6-1 (3)) button.
Lift moves downward (Laborota 4003 only); rotation, vacuum control and
timing start. The LED next to the Auto Start Stop (Fig. 6-1 (3)) button is lit.
If one function is already active, the remaining functions will be activated.
All distillation parameters can be changed while the Auto Start Stop function is
active.
1. Press Auto Start Stop (Fig. 6-1 (3)) button.
Lift moves upward (Laborota 4003 only); rotation, vacuum control and timing
stop. The LED next to the Auto Start Stop (Fig. 6-1 (3)) button goes out.
Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control
Using the Laborota 4002 / 4003 control
119

Publicité

Table des Matières
loading

Table des Matières