Manufacturer specification limits
Finnpipette F3 Variable Volume Single Channel
Range
Volume
µl
0.2-2 µl
2
0,2
1-10 µl, micro 10
1
1-10 µl
10
1
2-20 µl, micro 20
2
2-20 µl
20
2
5-50 µl, micro 50
5
5-50 µl
50
5
10-100 µl
100
10
20-200 µl
200
20
100-1000 µl
1000
100
0,5-5 ml
5000
500
1-10 ml
10000
1000
Finnpipette F3 Fixed Volume Single Channel
Fixed Volume
µl
1
5
10
20
25
50
100
200
250
500
1000
Adjustment
Adjustment is done only for one volume. The recommended adjustment volume is the minimum
volume or 10% of the maximum volume.
1. Place the service tool into the openings of the
calibration nut at the top of the handle.
2. Turn the service tool clockwise to increase, or
counterclockwise to decrease the volume.
3. After adjustment check the calibration according
to the instructions above.
Inaccuracy
µl
%
±0,050
±2,50
±0,024
±12,00
±0,100
±1,00
±0,025
±2,50
±0,100
±1,00
±0,035
±3,50
±0,20
±1,00
±0,06
±3,00
±0,20
±1,00
±0,06
±3,00
±0,30
±0,60
±0,15
±3,00
±0,30
±0,60
±0,15
±3,00
±0,80
±0,80
±0,30
±3,00
±1,2
±0,60
±0,36
±1,80
±6,0
±0,60
±1,0
±1,00
±25,0
±0,50
±10,0
±2,00
±50,0
±0,50
±20,0
±2,00
Inaccuracy
µl
%
±0,040
±4,00
±0,070
±1,40
±0,090
±0,90
±0,14
±0,70
±0,15
±0,60
±0,30
±0,60
±0,40
±0,40
±0,80
±0,40
±1,0
±0,40
±1,5
±0,30
±3,0
±0,30
Imprecision
s.d. µl
cv% calibration tip
0,040
2,00
0,020
10,00
0,050
0,50
0,020
2,00
0,080
0,80
0,030
3,00
0,08
0,40
0,05
2,50
0,08
0,40
0,05
2,50
0,15
0,30
0,125
2,50
0,15
0,30
0,125
2,50
0,20
0,20
0,10
1,00
0,4
0,20
0,14
0,70
2,0
0,20
0,6
0,60
10,0
0,20
4,0
0,80
20,0
0,20
8,0
0,80
Imprecision
s.d. µl
cv%
0,040
4,00
0,070
1,40
0,080
0,80
0,10
0,50
0,125
0,50
0,20
0,40
0,30
0,30
0,60
0,30
0,75
0,30
1,5
0,30
3,0
0,30
9
Factory
Flex 10
Flex 10
Flex 200
50
Flex 200
50
Flex 200
Flex 200
Flex 200
Flex 1000
5 ml
10 ml
Factory
calibration tip
Flex 10
Flex 10
Flex 200
Flex 200
Flex 200
Flex 200
Flex 200
Flex 200
Flex 1000
Flex 1000
Flex 1000