EN
8
Initial operation
Remove the transportation safety device from the bottom of the housing, see sheet "Transportation safety
device".
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specifications".
Turn on the mains switch. Switch position "".
The last used centrifuge data will be displayed.
Open the lid.
9
Opening and closing the lid
9.1
Opening the lid
The lid can only be opened when the centrifuge is switched on and the rotor is at rest. If it cannot be
opened under these circumstances, see the section on "Emergency release".
Swing handle rail on the lid upwards. The symbol " " (lid open) illuminates in the rotation indicator
Open the lid.
9.2
Closing the lid
Do not bang the lid shut.
Place the lid and swing handle rail on the lid downward. The symbol " " (lid closed) illuminates in the rotation
indicator
.
36/109
.