16
Instructions for Use
Use
Controller Controls, Indicators and Connectors
The following paragraph illustrates the Controller control panel and
interconnection panel. More details are contained in the appendix
"Technical Information".
Figure 2
1
Keyboard push-button for HIGH/LOW SPEED mode selection. It is active only when the
front panel operation has been selected. Pressed repeatedly, toggles between HIGH SPEED
and LOW SPEED.
2
Keyboard push-button for START, STOP, RESET mode selection. It is active only when the
front panel operation has been selected. By pressing once the starting phase begins; if
pressed again it stops the pump. If the pump has been stopped automatically by a fault, this
push-button must be pressed once to reset the controller and a second time to restart the
pump.
3
Keyboard push-button to recall on the display the cycle number, cycle time and pumplife
and serial number.
4
Keyboard push-button to recall on the display the pump current, pump temperature, pump
power, rotational speed, flow value detected, type of gas and alarm status. It is always
active regardless of the operating mode selected. Push-buttons 3 and 4, if pressed together
for at least 2 seconds put the controller in a routine where it is possible to select some
operation parameters.
5
LCD back-lighted alphanumeric display: dot matrix 2 lines x 16 characters.
198/262
5
Turbo V 2300
Turbo-V 3K-T
3
1
Controller 9699539, 869954 Front Panel
Turbo-V 2300 Rack Controller User Manual / 87-901-022-01 (B)
2
4