Technical Information
Use

Power Failure

In the event of a power failure (momentary or long term), the Turbo-
V 2300 controller will stop the turbopump and all the interconnected
pumps/devices. The Turbo-V 2300 vent valve, will vent the
turbopump only if the power failure is longer than the preset delay
time. When power is restored, the Turbo-V 2300 controller
automatically restarts the valves and the turbopump in the proper
sequence.
The display shows:
P
1
until normal operation achieved.

Remote Control Mode Operation

If remote signals are used to operate the controller, it must be
programmed for remote operation (see paragraph "Operating
parameter selections") and when ready to start, the display shows:
R
S
If the Soft Start has been deselected the display shows:
P
With or without Soft Start mode selected the START/STOP and LOW
SPEED front panel push- buttons are inoperative, while the CYCLE
NUMBER and PUMP CURRENT pushbuttons are always active.
254/262
U
M
P
I
2
X
X
X
E
A
D
Y
O
F
T
S
U
M
P
R
R
E
M
O
T
Turbo-V 2300 Rack Controller User Manual / 87-901-022-01 (B)
S
S
T
A
H
z
H
F
O
R
R
T
A
R
T
E
A
D
Y
:
E
S
T
A
R
T
I
N
G
J
H
S
E
M
O
T
E
H
S
U
S
E
R
T
H
S

Publicité

loading

Ce manuel est également adapté pour:

96995399699540