Instructions for Use
Installation
Installation
WARNING!
The Turbo-V controller is equipped with a 3-wire power cord and plug
(internationally approved) for user safety. Use this power cord and plug in
conjunction with a properly grounded power socket to avoid electrical shock.
High voltage developed in the controller can cause severe injury or death.
Before servicing the unit, disconnect the input power cable.
The Turbo-V controller can be used as a bench unit or a rack module, but it must
NOTE
be positioned so that free air can flow through the holes. Do not install or use
the controller in an environment exposed to atmospheric agents (rain, snow,
ice), dust, aggressive gases, or in explosive environments or those with a high
fire risk.
During operation, the following environmental conditions must be
respected:
To connect the controller to the pump use the cables kit 9699962. The
kit is not supplied with the controller. You have to connect both
cables contained in the kit to the connectors 4 and 8 (see picture on
next page) in order to run the pump. If one of the cables is not
connected the error "Check connection to pump" will be displayed.
See the appendix "Technical Information" for detailed Information
about the above mentioned and the other connections, and about the
options installation.
196/262
temperature: from O °C to +40 °C;
relative humidity: 0 – 95 % (without condensation).