INSTALLATION
!
High voltage developed in the controller can cause
severe injury or death. Before servicing the unit,
disconnect the input power cable.
The controller must be installed inside a rack module,
but it must be positioned so that free air can flow
through the holes. Do not install or use the controller in
an environment exposed to atmospheric agents (rain,
snow, ice), dust, aggressive gases, or in explosive
environments or those with a high fire risk.
During operation, the following environmental conditions
must be respected:
•
temperature: from 0 °C to +45 °C
•
relative humidity: 0 - 95%
(without condensation)
To connect the controller to the pump use the specific
cable supplied with the controller.
See the section "Technical Information" for detailed
information about the above mentioned and the other
connections, and about the options installation.
USE
This paragraph describes the fundamental operating
procedures.
Detailed
procedures that involve optional connections or options
are supplied in the paragraph "USE" of the section
"Technical Information".
Make all vacuum manifold and electrical connections
and refer to the pump instruction manual prior to
operating the controller.
WARNING!
NOTE
information
and
operating
Controller Controls, Indicators and
Connectors
The following paragraph illustrates either the front and
rear panel with its interconnections.
More details are contained in the appendix "Technical
Information".
Mini Vac
8
7
1. High Load LED, comes on when the current absorbed by the
pump reaches 12 mA
2. Output selected polarity LED
3. Output voltage scale (KV)
4. Main ON/OFF switch
5. Voltage/current LED scale
6. H.V. LED ON
7. Output connector of the sSignal proportional to the current
1 V= 1 mA log
8. Current scale (mA)