17
Technical Information

Use

Use

General

Make all vacuum manifold and electrical connections and refer to
Turbo-V 2300 pump instruction manual before to operating the
Turbo-V 2300 controller.
WARNING!
To avoid injury to personnel and damage to the equipment, if the pump is laying
on a table make sure it is steady. Never operate the Turbo-V 2300 pump if the
pump inlet is not connected to the system or blanked off.
The input signal P1 connector should be left in position including the shipping
NOTE
links if no external connections are made. The forepump and the Turbo-V 2300
pump can be switched on at the same time.
When the Turbo-V 2300 Rack Controller pump is baked by a membrane pump, the
NOTE
Soft Start mode should be deselected.
232/262
Turbo-V 2300 Rack Controller User Manual / 87-901-022-01 (B)