Connection J7 Purge-Vent

Figure 13
The integrated purge-vent valves can be driven by the controller.
Uuse the "purge-vent valves cable" delivered with the Turbo-V 2300
Rack Controller.
The Turbo-V 2300 pump integrates the .purge. and .vent valves. The
two valves are Normally Closed so if a power fail occur, the valves
will remain closed. The valves operating mode is independent from
the controller operating mode. The valve operating mode can be set
by serial line by win.125 or by front panel using the "program menu",
see following diagram for details.
Note that the vent and purge valve can also be controlled by means of
the serial connection (see the following diagram for details). The vent
valve connector is also used to connect the termistor placed on the
pump body to the controller. During normal operation, the motor is
fed with a 60 Vac and 530 Hz three-phase voltage. The power
provided to the motor is limited by the controller according to:
Turbo-V 2300 Rack Controller User Manual / 87-901-022-01 (B)
Connection J7 Purge-Vent
gas load (selectable by window 157: see the following "Window
Meanings" paragraph)
body temperature (can be read by window 210).
Technical Information
Interconnections
217/262