MESSAGE
OVERVOLTAGE
FLOW METER
ALARM
Turbo-V 2300 Rack Controller User Manual / 87-901-022-01 (B)
DESCRIPTION
A failure has occurred in the
controller power supply section,
or the controller has received a
spurious signal.
Alarm relating to the purge flow.
The flow value remained at less
than the set threshold for 10
seconds or more.
Instructions for Use
Error Messages
REPAIR ACTION
Press the START button twice to start
the pump again. If the message is
redisplayed, contact Agilent for
maintenance.
Check the correct operation of the gas
supply circuit.
16
203/262