Symbols on the unit and product label
CE marking
f
Date of manufacture
g
Manufacturer
h
Max. load
j
I nstructions: Read the
instructions and make sure
you understand them
before using this product
Article number/reference
l
number
Serial number
n
Functions
Emergency stop
In an emergency, the emergency stop can be
activated by pulling the red belt. To reset the
emergency stop, push up the activation plate
at the top of the belt.
Emergency lowering
In the event of an emergency, the emergency
lowering button on the lift or hand control can
be pressed to quickly lower the lift. An audio
signal indicates that the lift is lowering. Always
ensure that emergency lowering can be per-
formed as safely as possible for the patient. The
emergency lowering button only works when
the lift is switched on and emergency stop has
been deactivated. When the emergency lower-
ing is active, the end position switches are inac-
tive. If the button is pushed until the lift is in its
lowest position, the lift will start to go up again.
Always stop at the lowest position.
The emergency lowering button
o
must only be used in emergency
situations.
Safety catch
The lift is equipped with a safety catch. This
prevents involuntary falls. If the lift belt is
www.humancaregroup.com
Warning
o
The lift contains lead bat-
q
teries and electronics and it
must be recycled appropri-
ately.
Third party certification
marking (not applicable for
all regions and versions)
Type-B product
p
Switch the product off/
Switch the product on
pulled out too quickly, the safety catch will
engage, thus preventing the fall.
Mechanical emergency lowering
During power loss and the hoist is loaded, user
in the hoist, the mechanical emergency lowering
is activated and the hoist is lowering in a safe
and reliable manner.
Overheating
The lift has overheating protection which stops
the motor if it gets too hot. Overheating can
occur if the motor is overloaded or operated
continuously for quite a long time. The lift's work
cycle (work:pause) is 15:85. Maximum continu-
ous operation is two minutes. If this is exceeded,
there is a risk of the motor overheating.
Belt tension guard
The lift is equipped with a belt tension guard
that prevents unintentional discharging of
the lift belt. It must be loaded in order to be
able to pull the belt out. Pull gently on the
belt while at the same time pressing the Down
button on the hand control or control panel.
The weight of the bar is sufficient for the belt
to be fed out.
ENG
ENG
25