Maintenance
13
Maintenance
13.1 Maintenance intervals
The device should be inspected at least annually by suitably
qualified persons.
Charge the lead battery after each application, at the latest
however after the battery alarm has been triggered.
Maintenance by suitably qualified personnel – every year.
The inspection intervals must be established in each
individual case and shortened if necessary, depending on
technical safety considerations, engineering conditions,
and the technical requirements of the equipment.
We recommend that a service agreement be concluded
with Dräger and that repairs also be carried out by
DrägerService.
NOTICE
i
i
For the gas detection instruments used, the
maintenance intervals specified in the relevant
Instructions for Use apply.
13.2 Replacing the filter
NOTICE
i
i
The filter can only be replaced during pump operation
or operation with X-am 5100.
13.2.1 Replacing the pump filter
1. Unscrew the connecting bush from the filter.
2. Unfasten the lock at the holder (pump).
3. Remove the holder (pump).
4. Unscrew the filter anti-
clockwise.
5. Screw a new filter (order
no. 83 19 364) onto the
device in a clockwise
direction.
6. Place the holder (pump)
onto the housing.
7. Fasten the lock at the
holder (pump).
8. Screw
the
connecting
bush onto the new filter.
64
13.2.2 Replacing the gas inlet filter (only for operation
with X-am 5100)
NOTICE
i
i
Dräger recommends to replace the gas inlet filter
regularly depending on the dust exposure.
1. Unscrew the gas inlet filter
anti-clockwise.
2. Screw a new gas inlet
filter (order no. 83 23 016)
onto the device in a
clockwise direction.
13.3 Replacing the batteries
WARNING
!
Explosion hazard!
Do not replace the batteries in areas subject to
explosion hazard! Batteries are part of the Ex
approval.
Only the following types may be used:
Battery pack, 24 Ah (PBT 00X1), order no. 83 22 921
1. Switch off the device (see Chapter 3.2 on Page 42).
2. Unfasten the screws (M5
cylinder-head screw with
internal hexagon) on the
bottom
side
housing.
3. Lift the upper part of the
housing and disconnect
the plug connection from
the bottom plate.
4. Release the four M5 nuts.
5. Disconnect
connections
01433112.eps
bottom plate.
6. Replace the old battery
block with a new one.
7. Re-establish
connection to the bottom
plate.
8. Check correct fit of the
O-ring.
9. Tighten the four M5 nuts.
of
the
the
cable
from
the
the
cable
Dräger X-zone 5500 / 5800
05733112.eps
01533112.eps
01633112.eps