OPERATION: How to switch from one weighing range to the other and how to calibrate the balance.
Switching from one weighing range into the other is done by using the single control bar.
Before working with the balance for the first time, it must be calibrated.
Calibration means setting the balance to the correct weight display.
The balance is equipped with a built-in calibration weight (for specifications, see "Technical specifications").
NOTE: to achieve stable results, it is recommended that the balance be left connected to the power supply for 60 minutes prior to calibrating.
Irrespective of the selected weighing range, the calibration procedure applies to both weighing ranges at the same time.
rng
rng
Switching weighing ranges
– Press single control bar (1) until the balance
displays rng. (rng = range).
– Releasing and then briefly pressing the control
bar again, allows you to switch between the
30 g range and the 160 g weighing range.
When the weighing range wished has been selected,
wait: Display "- - - -", then "0.0000".
The balance is now in the weighing mode.
30
160
Preparations for calibrating the balance
– Remove all objects from the pan.
– Close the sliding windows.
Set the balance to the calibration mode:
– Press control bar (1) until "CAL" is displayed.
– Release control bar.
"CAL - - - -", will appear in the display.
CAL
----
CAL
100
Calibrating the balance
The calibration weight is placed in position by means
of the calibration lever (3).
As soon as "CAL 100" appears (the 100 blinks),
slowly move the calibration lever towards the rear
all the way to the stop.
First, "CAL - - - -" appears, then "100.0000". When
the display "CAL 0" appears (0 blinks), move lever
back to original position. Wait: display "- - - -"
followed by "0.0000" appears.
The balance is now calibrated in both weighing
ranges.
7