OPERATION: How to calibrate the balance.
Before working with the balance for the first time, it must be calibrated.
Calibration means setting the balance to the correct weight display.
The balance is equipped with a built-in calibration weight (for specifications, see "Technical specifications").
NOTE: to achieve stable results, it is recommended that the balance be left connected to the power supply for 60 minutes prior to calibrating.
CAL
CAL
Preparations for calibrating the balance
– Remove all objects from the pan.
– Close the sliding windows.
Set the balance to the calibration mode:
– Press control bar (1) until "CAL" is displayed.
– Release control bar.
"CAL - - - -", will appear in the display.
----
100
Calibrating the balance
The calibration weight is placed in position by means
of the calibration lever (3).
As soon as "CAL 100" appears (the 100 blinks),
slowly move the calibration lever towards the rear.
First, "CAL - - - -" appears, then "100.0000". When
the display "CAL 0" appears (0 blinks), move lever
back to original position. Wait: display "- - - -"
followed by "0.0000" appears.