Technical Documentation (Con't)
A
IEC 60601-1-2, Table 206
Recommended separation distance between portable
and mobile RF communications equipment and Basic 30
Basic 30 is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of Basic 30 can help prevent electromag-
netic interference by maintaining a minimum distance between portable and mobile RF com-
munications equipment (transmitters) and Basic 30 as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
W
150 kHz to 80 MHz
d = 1.2 √ P
0.01
0.12
0.38
0.1
1.2
1
3.8
10
12
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the trans-
mitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3 √ P
0.23
0.73
2.3
7.3
23
145