8. Read the outlet pressure on the gauge and, if necessary, adjust by turning the handwheel.
9. Do not leave the pressure regulator and the patient unsupervised during
GB
To increase the outlet pressure
5.1.2 Non-adjustable pressure regulators
1. Check on the type plate whether the appliance is suitable for the intended
2. Prepare the downstream equipment and accessories in accordance with the
3. Appliances with NIST inlet connection: Connect the TESCOM EUROPE hose with inlet
connection in accordance with the relevant instructions for use.
4. Remove the protective caps covering the connections on the appliance and keep them
in a safe place.
5. Connect the inlet connection on the pressure regulator to the gas extraction point.
The temperature of the gas and the pressure regulator drops during operation. This
can produce hypothermia in the patient.
Observe the minimum operating temperature of the downstream equipment and
accessories. The temperature of the gas and the pressure regulator drops during operation.
5.2

Decommissioning and storage

5.2.1 adjustable pressure regulators
Turn the handwheel on the pressure regulator as far as it will go to the left (anti-clockwise)
and allow the gas in the pressure regulator to escape.
operation.
purpose (type of gas, pressure). The maximum permissible inlet pressure of
the appliance must be greater than or equal to the line pressure.
relevant instructions for use and use the correct and proper procedures to
connect them to the outlet connection on the appliance.
The connections must match. Do not use adapters nor modify the connections
on the appliances.
To reduce the outlet pressure (only while gas is flowing)
26