Sealing Plug
O
Grey adapter
Double tube
Sealing Plug
5. Rinse the creased tube thoroughly inside and out with clear warm water.
6. Thoroughly shake out the tube system.
7. Hang up the tube system and allow it to drip dry, in order to prevent moisture entering
your therapy device.
8. Use the therapy device to dry the tube system completely. For details see the instructions
for use for your therapy device.
Disinfecting the housing
The housing, the connecting sockets and the connecting cable can be cleaned by simply
wiping them with disinfectant. We recommend terralin
Disinfect the tube system
As a disinfectant we recommend GIGASEPT FF. You are recommended to wear suitable
gloves (e.g. household or disposable gloves) during disinfection procedures. Observe the
instructions for the disinfectant used.
No disinfectant may penetrate the double tube.
To disinfect, proceed as follows:
Remove double tube:
Grey adapter
Double tube
Sleeve
Transparent adapter
38
EN
Hygienic preparation
1. Detach the tube system from the device and from the
supply tube
2
exhalation system.
2. Pull the double tube a little way out of the creased tube.
Seal the opening with the sealing plug supplied.
3. Close the small opening of the grey adapter and the O
supply tube with the sealing plugs supplied.
4. Clean the creased tube thoroughly to remove all resi-
dues, using warm water containing a little washing-up
liquid. Thoroughly wash out the inside of the tube.
1. Release both sleeves from the adapters.
2. Pull the double tube out of the creased tube.
3. Take the short piece of the double tube off the gray
adapter. Then take the tube out of the transparent
adapter.
Sleeve
4. Release the long part of the double tube from the
transparent adapter.
®
protect.
2