Transport and storage / Assembly
4
Storage
Storage and transport temperature
Humidity
Shelf life of sensor and electrolytes
in original packaging
Contents
• 1 CLE sensor with membrane cap and tensioning ring
• 1 bottle electrolyte (100 ml)
• 1 set of operating instructions
• 1 screwdriver
5
Pouring
Electrolyte
22
Transport and storage
NOTE
Transport, send and store the sensor in the original
packaging. Retain the packaging complete with
styropor parts.
NOTE
When sensor is stored for longer than the shelf life period,
send it to ProMinent for checking or reconditioning. Safe
function and accuracy of measurement cannot otherwise
be guaranteed.
Assembly
IMPORTANT
Do not touch, damage or bring into contact with greasy
substances the white membrane or the electrodes on the
electrode shaft. The sensor will not, in such cases, work
accurately.
Replace the membrane cap or send the sensor to ProMinent
to have the electrodes cleaned.
NOTE
Carry out the following actions over a washbasin.
Remove the red cap completely from the nozzle and cut the nozzle at the
marked position to open the nozzle canal.
Remove the membrane cap cover and unscrew the membrane cap from
the electrode shaft.
Rinse the membrane cap and the electrode with a little electrolyte.
Fill the membrane cap up to the rim with electrolyte.
Remove air bubbles by lightly tapping the membrane cap on an even
surface.
5 - 50 °C
max. 90 % rel. humidity,
non condensing