Options de sortie de données
SIWXS...-3-..-...
or
ISX...-..-...
Option A12
RS422
TxD+
TxD-
RxD+
RxD-
DTR+
DTR-
CTS+
CTS-
Signal_GND
Important note!
Max. voltage V
of any electrical apparatus connected to the interface converter type YDI05-Z.. may not
m
exceed 250V. The internal Zener voltage V
Note:
1: External power supply (e.g. type 6971991 in the scope of delivery).
2: DC power supply: 12 to 30 V; positive voltage on Pin 1 of LV4; order cable and cable gland separately
2013-02-02
Dr. D. Klausgrete
44
ISXBBS
Hazardous area
G
F
M
E
J
N
C
K
A
B
up to 1000m
14-
contact
N
green
K
yellow
G
pink
M
gray
C
violet
B
black
F
red
E
blue
A,J
brown, white
is 12.6V.
z
RS-422 (EX) - RS-232 (Non-EX) connection
Non-hazardous area
DC power supply
12V DC (min) to
30V DC (max)
LV3
YDI05-Z
interface
converter
(see Sheet 4)
LV3
9
RxD+
10
RxD-
7
TxD+
8
TxD-
5
CTS+
6
CTS-
3
DTR+
4
DTR-
1,2
Title
Drawing number
1)
LV4
or
DC
RS-232
connection:
LV1
see page 1
1
6
12-
LV1
contact
or
4
B
TxD
1
C
RxD
3
D
DTR
2
H
CTS
5
E
Signal_GND
Data Transfer
3
Page
of
00
36953-751-40
Revision
2)
6