Correction of the Operating Manual:
For the following commands, the correct *RST value is given below:
:CONFigure[:BTS]:NETWork:PHASe
:CONFigure[:MS]:NETWork:PHASe
These commands select the phase of the standard according to which the base/mobile station will
work.
Example:
Features:
Mode:
Extended Functionality:
:[SENSe<1|2>:]AVERage:TYPE MAXimum | MINimum | SCALar | VIDeo | LINear
This command selects the trace averaging method.
VIDeo
LINear
Example:
Features:
Mode:
Note:
It is also possible to select the evaluation mode (MAXimum, MINimum, SCAlar) for the
trace with this command. However, it is recommended to use command
DISPlay[:WINDow<1|2>]:TRACe<1...4>:MODE for this purpose. The command
AVERage:TYPE should be used only to select the trace averaging method. Also, the
query reads out the trace averaging mode only.
The following functions are defined but should not be used:
MAXimum (MAX HOLD):
MINimum (MIN HOLD):
SCALar (AVERAGE):
1065.6000.30
":CONF:NETW:PHAS 2"
*RST value:
2
2,PLUS
SCPI:
device-specific
BTS/MS
Averaging of logarithmic level values.
Averaging of linear power values prior to their conversion into level values.
":AVER:TYPE LIN"
*RST value:
VIDeo
SCPI:
device-specific
A, VA, BTS, MS ("VIDeo" and "LINear" are not available in VA mode)
for PGSM, EGSM, GSM1800
for RGSM
for GSM850 and GSM1900 no phases are available
AVG(n) = MAX(X 1 ...X n )
AVG(n) = MIN(X 1 ...X n )
n
1
×
AVG n
( )=
xi
n
=
J