Technical Data - Soehnle CONNECT 400 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 34

TECHNICAL DATA

Item number
Power supply
EN
Display mode
Measurement mode
Measurement range
Accuracy
Normal operating
conditions
Storage and transport
conditions
Measurement perime-
ter on the upper arm
Weight
External dimensions
Accessories
Operating mode
Protection category
Water ingress
protection
Device Classification
Software version
Bluetooth
30
68097
Battery operation: 4 alkaline batteries AAA 6V DC
Power adapter operation 6 V 1 A
(Please use only the recommended power adapter.)
LCD display (100 x 68 mm / 3.94 × 2.68 inch)
Oscillographic testing mode
Target cuff pressure: 0 mmHg – 300 mmHg (0 kPa – 40 kPa) measurement
pressure
SYS: 60 mmHg – 230 mmHg (8.0 kPa – 30.7 kPa)
DIA: 40 mmHg – 130 mmHg (5.3 kPa – 17.3 kPa)
Pulse: (40 – 199) beats per minute
Pressure: 5 °C – 40 °C within ± 3 mmHg (0.4 kPa) pulse at: ± 5 %
Temperature: 5 °C to 40 °C
Relative humidity: ≤85 % RH
Atmospheric pressure: 86 kPa to 106 kPa
Temperature:-20 °C to 60 °C
Relative Humidity: 10 % RH-93 % RH
Atmospheric Pressure: 50 kPa-106 kPa
22 – 42 cm / 8.66 – 16.54 inch
approx. 277 g / 9.7 oz (excluding batteries and cuff)
approx. 154 ×106 × 57 mm / 6.06 × 4.17 × 2.24 inch
4 alkaline batteries AAA
Continuous operation
Applied part type BF
IP21 means that the blood pressure monitor is protected against solid foreign
bodies with a size of 12.5 mm / 0.49 inch and larger as well as vertically
dripping water.
Battery Powered Mode: Internally Powered ME Equipment
AC Adaptor Powerde Mode: Class II ME Equipment
V01
Bluetooth Module No.: AW 2540 MV 1
High frequency range: 2402 MHz to 2480 MHz
Output power level: -1 dBm
Supply Voltage: 2 V - 3.6 V
Transmitting Distance: 10 meters
WARNING: Modifications of the device are not permitted.
(Modell: TMB-1585-BT)