INTRODUCTION
by the manufacturer. Otherwise,
danger may arise for the user and
EN
the device.
• Please inform Soehnle customer
service of unexpected operating
status or events.
• The optional adapter is a compo-
nent of the MEDICAL ELECTRICAL
DEVICE.
• The plugs/pins of the adapter
plug isolate the device against the
mains power supply.
• Please do not place the device in
a location in which it is difficult to
disconnect from the power supply.
• Keep the device outside of the
reach of small children/pets to
prevent the inhalation or swallo-
wing of small parts. The cable/cord
poses a risk of strangulation.
• When stored at the minimum sto-
rage temperature between uses,
electrical medical devices must be
warmed up for at least 30 minutes
before they are operational. When
stored at the maximum permissib-
le storage temperature between
uses, electrical medical devices
must cool off for at least 30 minu-
tes before they are operational.
• Please carefully monitor a possible
interruption of blood circulation
and a possible injury of the patient.
• Do not place the cuff over a
wound as this may cause further
injuries.
• Do not place the cuff on the same
arm on which other electrical
medical devices are placed.
20
• Please ensure that the operation
of the device does not interrupt
blood circulation for a longer
period.
• Avoid squeezing or constriction of
the connecting tube.
• The Blood Pressure Monitor
has been clinically tested pursuant
to the requirements of
EN 1060-4:2004.
Metrological maintenance and
care information
• Do not perform any metrological
maintenance if the device is in use.
• For reliable operation, we recom-
mend that the device is subject to
metrological control once every
two years pursuant to Sec. 14
of the Medical Product Operat-
ing Directive (MPBetriebV). We
recommend testing the device's
performance once every two years
and after metrological control and
repair, at a minimum by testing
the tolerance requirements of the
pressure display and tightness
of the cuff (testing at least at 50
mmHg and 200 mmHg).
• The metrological control must be
performed by a licensed tester or
testing institute.
• We recommend cleaning the de-
vice and cuff only with a soft cloth.
Storage information
• If the device is not being used,
store it and the adapter in a dry lo-
cation away from heat, dust, fibres,
lint, and out of direct sunlight. Do
not place heavy objects on the
storage box.