COMPACT PIEZO LED
CAUTION: In the case that the end
user, when operating in his or her own
medical study or clinic, must subject the
electro-medical equipment and systems to
periodical inspections in order to adhere to
imposed requirements, the test procedures
EN
that must be applied to electro-medical
equipment and systems to evaluate safety
must be performed in line with norm EN
62353 'Electro-medical devices - Periodical
inspections and tests to be performed after
repair interventions on electro-medical
devices'.
WARNING: Checking device status
before the treatment. Always check that
there is no water underneath the device.
Before every treatment, always check
that the device works perfectly and that
the accessories are efficient. In case you
uncover operating abnormalities, do
not perform the treatment. Contact an
Authorized Mectron Service Center if the
abnormalities concern the device.
WARNING: Cleaning and sterilization
of the new or repaired parts. All new and
repaired accessories are supplied in non-
sterile conditions. Before use, and after
each treatment, they must be cleaned and
sterilised in strict compliance with the
instructions given in the Manual of Use
and Maintenance, Chapter 06 - CLEANING
AND STERILIZATION.
4
CAUTION: The electrical contacts inside
the cord and handpiece connectors must
be dry. Before connecting the handpiece
to the cord, make sure that the electrical
contacts of both connectors are perfectly
dry, especially after the sterilization cycle
in autoclave. If need be, dry the contacts by
blowing compressed air onto them.
CAUTION: Do not activate the
handpiece while the insert is in contact
with the part to be treated. Doing so, will
not allow the electronic control circuit
of the device to recognize the best point
of resonance of the insert, required for
efficient and optimum performance.
WARNING: Only use original Mectron
inserts, accessories, and spare parts.
WARNING: Use of non-original
Mectron inserts: this use entails finite
damage to the handpiece threading, thus
compromising correct operation and
risking to cause harm to the patient.
WARNING: Do not change the insert
with handpiece running, to prevent
causing wounds to the operator.
CAUTION: Always make sure to read
the Use and Maintenance Manual of the
compact piezo LED for all information
on use, cleaning and maintenace of the
device.