COMPACT PIEZO LED
Problem
EN
No liquid outflows
from the insert during
operation
Insufficient
performance
NOTE: Always refer to the Use and
Maintenance Manual supplied
with the compact piezo LED for all
information on sending the device to
The indications that appear in this publication are not binding and can be modified without fore-notice.
The Italian version of this manual is the original document from which its translations have been obtained. In case of any
discrepancy, the Italian version will have pertinence.
Texts, images, and graphics of this manual are property of Mectron S.p.A., Carasco, Italy. All rights reserved.
The contents cannot be copied, distributed, changed, or made available to third parties without the written approval of
Mectron S.p.A.
20
Possible cause
The insert type does not foresee
liquid passage
The insert is obstructed
The handpiece is obstructed
The Water function is off
The insert is not correctly
tightened on the handpiece
Insert broken, worn-out, or
deformed
an Authorized Mectron Service Centre
and on the warranty terms of the
product.
Solution
Use an insert type with liquid
passage
Unscrew the insert from the
handpiece and free the insert
water passage by blowing
compressed through it.
If the problem persists, replace
the insert with a new one
Contact an Authorized Mectron
Service Center
Turn on the water function on
the dental chair
Unscrew and correctly screw the
insert again with the Mectron
torque wrench (see Use and
Maintenance Manual compact
piezo LED- paragraph 04.4
INSTRUCTION FOR USE)
Replace the insert with a new
one