05 Û DISPOSAL MODES AND PRECAUTIONS
WARNING: Hospital wastes.
Hospital wastes. Treat the following items
as hospital waste:
- Inserts: when they are worn out or
broken;
- Torque wrench for inserts: when worn
out or broken.
The use and throw-away materials and
materials that entail a biological risk must
be disposed of in accordance to the norms in
06 Û TECHNICAL DATA
Device compliant to Dir. 93/42/CEE:
Classification as per EN 60601-1:
Device for intermittent operation:
Power supply voltage:
Max. power absorbed:
Fuse:
Output control electrovalve:
Operating frequency:
Power types:
Water supply:
LED system of the handpiece:
Protections of the APC circuit:
Operating conditions:
Transport and storage conditions:
05 Û DISPOSAL MODES AND PRECAUTIONS
force locally concerning hospital wastes.
compact piezo LED must be disposed of and
treated as a waste for separate collection.
Disregard of the previous points may entail a
fine, pursuant to Directive 2002/96/CE.
It is up to the purchaser to hand over
the device for its disposal to the retailer
who supplies him new equipment; the
instructions for proper disposal are available
from Mectron.
Class IIa
The class definition is delegated to the dental chair
manufacturer using the compact piezo LED in its unit
Applied part type "B" (handpiece, insert)
60sec. ON - 30sec. OFF with irrigation
30sec. ON - 120sec. OFF without irrigation
Power supply compliant to IEC/EN 60601-1:
24V
50/60 Hz, or
32V
40 VA
T2AL, 250V (not included)
24V
Automatic scan
From 24 KHz a 36 KHz
Adjustable according to the instructions given by the
dental chair manufacturer
Adjustable according to the instructions given by the
dental chair manufacturer
Operating pressure from 1 to 6 bar
Adjustable according to the instructions given by the
dental chair manufacturer
No handpiece detected
Cord interruption
Insert not tightened correctly or broken
from +10°C to +40°C
Relative humidity from 30% to 75%
Pressure of air P: 800hPa/1060hPa
from -10°C to +70°C
Relative humidity from 10% to 90%
Pressure of air P: 500hPa/1060hPa
2,5W Max.