Changing Reusable Jar - Medela Vario Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 70

8 Changing reusable jar

18 |
Changing reusable jar
1.
or
2.
3.
4. Empty jar and dispose of patient tubing in accordance with
local guidelines and inhouse guidelines.
5. Switch on the Vario. The pump will start running.
6. Set vacuum according to the particular application.
See chapter 7.3.
1.1 Switch the on / off switch
of the Vario to position «0»
(c/i - versions: «Off»).
1.2 Remove patient and
vacuum tubings from the
lid.
2.1 Remove full jar from the
Vario.
3.1 Prepare new jar and
unpack it.
3.2 Secure new jar to the
Vario
3.3 Reconnect vacuum tubing
and new patient tubing
firmly.

Publicité

Table des Matières
loading

Table des Matières