6 Preparation for use

WARNINGS
For use only by medically trained persons who have been adequately trained in suction
procedures and in the use of aspirators.
CAUTIONS
– The Vario must remain in an upright position during use.
– Sterile accessories must be checked on the integrity of the packaging before use.
– Non sterile and reusable accessories must be cleaned, disinfected and / or sterilized
according to the Medela cleaning guide (product code 200.2391).
6.1

Checks before use

– Check the Vario system before use for damage of the power cord or plug,
obvious device damage or safety defects and proper functioning of the device.
– Check for completeness and general condition of the Vario delivery package.
– Only for AC / DC versions: make sure that the rechargeable battery is calibrated
according to chapter 16.
– Check all accessories prior to use:
– suction jars, lids and liners for cracks, brittle and flawed spots. Replace if
necessary.
– Tubing for cracks, brittle areas and that connectors are firmly attached.
Replace if necessary.
– As an additional safety test, evacuate the system (including jars) to maximum
vacuum several times before actual use.
6.2
Assembly of the
basic configuration
1.
2.
Attach all necessary accessories according to your needs.
See annex "II List of accessories".
1.1 Attach the overflow
protection bacteria filter to
the Vario with the arrow
pointing in the flow
direction.
Preparation for use