Télécharger Imprimer la page

Bühler technologies Baseline Notice De Montage page 4

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Baseline
1 Introduction
This quick guide will assist you in starting up the unit. Follow
the safety notices or injury to health or property damage may
occur. Carefully read the original operating instructions in-
cluding information on maintenance and troubleshooting
prior to startup. These are located on the included CD and on-
line at
www.buehler-technologies.com
Please direct any questions to:
Bühler Technologies GmbH
Harkortstraße 29
40880 Ratingen
Germany
Tel.: +49 (0) 21 02 / 49 89-0
Fax: +49 (0) 21 02 / 49 89-20
These operating instructions are a part of the equipment. The
manufacturer reserves the right to change performance-, spe-
cification- or technical data without prior notice. Please keep
these instructions for future reference.
1.1 Intended Use
The portable sample gas probe (probe) is intended for use in
gas analysis systems in industrial applications. It is particu-
larly suitable for use with mobile sample gas conditioning.
Portable sample gas probes are among the main components
in a gas conditioning system.
• Therefore also note the related drawing in the appendix
or in the data sheet on the CD.
• Before using the device, verify the listed technical data
meet the application parameters.
• Further check if all contents are complete.
Please refer to the nameplate to identify your model. In addi-
tion to the job number it also contains the item number and
model designation.
When connecting, please note the specific values of the
device, and the correct version when ordering spare parts.
1.2 Contents
• 1 x Sample gas probe
• Connection- and mounting accessories per delivery note
• Product documentation (brief instruction + CD)
2 General hazard warnings
Installation of the device shall be performed by trained staff
only, familiar with the safety requirements and risks.
Check all relevant safety regulations and technical indications
for the specific installation place. Prevent failures and protect
persons against injuries and the device against damage.
The operator of the system must secure that:
• safety and operation instructions are accessible and fol-
lowed,
• local safety regulations and standards are obeyed,
• performance data and installation specifications are re-
garded,
• safety devices are installed and recommended mainten-
ance is performed,
4
Bühler Technologies GmbH
• national regulations for disposal of electrical equipment
are obeyed.
DANGER
Toxic, corrosive gases
The measuring gas led through the equipment can be hazard-
ous when breathing or touching it.
a) Check tightness of the measuring system before putting
it into operation.
b) Take care that harmful gases are exhausted to a save
place.
c) Before maintenance turn off the gas supply and make
sure that it cannot be turned on unintentionally.
d) Protect yourself during maintenance against toxic / cor-
rosive gases. Use suitable protective equipment.
3 Transport and storage
The product should only be transported inside the original
packaging or a suitable alternative.
When not in use, the equipment must be protected from
moisture and heat. They must be stored in a covered, dry and
dust-free room at a temperature between -20 °C and 50 °C.
4 Installation and connection
4.1 Installation site requirements
Portable sample gas probes are intended as temporary gas
sampling. Optional assembly plug and flanges allow for safe
and gas-tight installation.
• Installation site and installation position are determined
based on requirements specific to the application.
• If necessary, the connection piece should be slightly tilted
toward the centre of the channel.
• Also ensure adequate and safe access for installation. Par-
ticularly consider the available installation length for the
built-in sampling tube and, if applicable, the intake filter.
If the probe is transported to the installation site in pieces, it
will first need to be assembled.
4.2 Installation
WARNING
Gas leakage, condensate, hot surfaces
Burns and / or health hazards due to gas / condensate!
a) The sample gas and/or condensate may be hazardous to
the health.
b) Parts near the sampling point may be hot.
c) The probe may become very hot due to the process
gasses.
ð Close the gas supply, if necessary flush the probe with air,
and allow the probe to cool down before beginning in-
stallation or maintenance.
ð Wear safety gloves and goggles.
The portable probe can be installed in the site using the avail-
able mounting accessories such as cones, plugs, and flanges.
Attach the suitable mounting element and, if applicable, the
intake filter (optional) to the sampling tube for the probe.
BX464001 ◦ 09/2016

Publicité

loading