Schiller Argus VCM Mode D'emploi page 97

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

SpO2/Pulse Rate
Range
Low Perfusion
Accuracy
Range
Accuracy
Standards
1
Neonate specifications are shown for neonate sensors with ARGUS VCM. Saturation accuracy will vary
by sensor type recommended by the manufacturer..
Specification applies to monitor performance and was validated with Biotek and Nellcor simulators
2
Temperature
Probe Type
Range
Display Accuracy
Measurement units
Measurement modes
Predictive Mode
Monitored Mode
Standards
Printer
Type
Resolution
Printing speed
Paper width
%Saturation
1% to 100%
0.03% to 20%
Without Motion-Adults 70% to 100% ±2 digits
1% to 69% unspecified
Without Motion-Neonate
1
70% to 100% ±3 digits
1% to 69% unspecified
Low Perfusion
70% to 100% ±2 digits
2
1% to 69% unspecified
Pulse Rate
20 BPM to 300 BPM
Without Motion
20 BPM to 300 BPM ±3 digits
2
Low Perfusion
20 BPM to 300 BPM ±3 digits
2
EN865:1997
Thermistor probe
26° C to 43° C (80° F to 110°F)
±0.1° C (±0.2° F)
° C, ° F
Predictive, Monitored
One-time measurement in a single temperature reading which
is displayed at the end of the brief measurement period
Continuous measurement over an indefinite period.
ASTM E1112-00, EN12470-3
Thermal
8 (dots/mm)
45 (mm/s)
57 (mm)
Operator's Manual
ARGUS VCM
Specifications
91

Publicité

Table des Matières
loading

Table des Matières