Battery Operation
The monitor cannot operate with a fully discharged battery. Before turning on
which battery has been completely discharged, first plug the monitor into an AC outlet to
charge the battery for a few minutes. The monitor may then be powered on.
A new, fully charged battery will provide 2 hour of monitoring time under the following
conditions:

Charging a Low Battery

1. Connect the monitor to AC power in order to charge a low or dead battery.
(see the Setting up the Monitor section)
Note: A full charge of a dead battery takes approximately 12 hours while the monitor is
turned off.

Low Battery Indication

Battery Indicator is flashing when the remaining battery power is only enough for about 10
minutes of operation. The monitor will also sound an audible alarm. This will inhibit the
monitor from taking printing. This audible alarm can be silenced by pressing the Alarm
silence button. Connecting the monitor to AC power will terminate the alarm.
The monitor remaining the battery status of only about 3-minute operation will inhibit the
monitor from taking NIBP inflation.
After a critical low battery alarm sounds for about 5 seconds,
automatically shut down. You should connect the monitor to an AC power source to avoid
any loss of monitor settings and trend data
22
ARGUS VCM
No audible alarms sound
No serial output devices are attached to
No printing
All the monitoring parameters are active with one NIBP measurement
per 15 minutes
Operator's Manual
ARGUS VCM
ARGUS VCM
ARGUS VCM
monitor will
ARGUS VCM