Factory Defaults
General
Parameter Ranges and Default Settings
General
is shipped with factory default settings. You may change the default settings to
ARGUS VCM
desired values as described in 'Using the Monitor' and 'Alarm and Limits' sections of this
manual. When you turn the monitor off properly with the change of the default setting, the
changed settings will be saved and they will be power-on default settings for next power-on.
If you want the power-on setting back to factory defaults, refer to page 32 'Using the
Monitor' section of this manual.
Parameter Ranges and Default Settings
Parameter
Systolic (mmHg)
Alarm Limits
Diastolic (mmHg)
Alarm Limits
MAP (mmHg)
Alarm Limits
PR (bpm)
Alarm Limits
SpO2 %
Alarm Limits
Patient Type
NIBP Units
NIBP Auto Interval
Temperature Units/Modes
Pulse Tone Volume
Alarm Volume
Night Panel
Print Control
Table 11. Parameter Ranges and Factory Defaults
Ranges
Neonatal
Low: 25 to 115
Pediatric
Low: 30 to 155
Adult
Low: 30 to 255
Neonatal
Low: 10 to 100
Pediatric
Low: 15 to 125
Adult
Low: 20 to 250
Neonatal
Low: 10 to 105
Pediatric
Low: 15 to 135
Adult
Low: 20 to 250
Neonatal
Low: 25 to 295
Pediatric
Low: 25 to 295
Adult
Low: 25 to 295
Neonatal
Low: 50 to 98
Pediatric
Low: 50 to 98
Adult
Low: 50 to 98
Off, STAT, 1, 2, 3, 4, 5, 10, 15, 30, 45, 60, 90, 120,
ARGUS VCM
High: 30 to 120
High: 35 to 160
High: 35 to 260
High: 15 to 105
High: 20 to 130
High: 25 to 255
High: 15 to 110
High: 20 to 140
High: 25 to 255
High: 30 to 300
High: 30 to 300
High: 30 to 300
High: 52 to 100
High: 52 to 100
High: 52 to 100
Adult, Pediatric, Neonatal
mmHg, kPa
240
°C, °F, °C M, °F M
0 to 8
1 to 8
On, Off
Manual, Stream
Operator's Manual
Defaults
50, 100
75, 145
75, 220
30, 70
35, 100
35, 110
35, 80
50, 110
50, 120
100, 200
50, 150
50, 120
85, 98
90, 100
90, 100
Adult
mmHg
15
°F
4
4
Off
Manual
87