Accessory/Replacement Items; Set-Up; How To Operate Your Suction Unit; Battery Operation - Laerdal 88 00 20 Guide D'instructions

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 16
E n g l i s h
4

ACCESSORY/REPLACEMENT ITEMS

The following items can be purchased separately as accessories or replacement items for your suction unit:
Description
Switch mode AC to DC adapter/charger
Line cord for US
Hospital grade line cord (120 VAC)
Line cord for UK
Line cord for Continental Europe
Line cord for Australia
12V DC power cord (1 each)
Fuse for 12VDC Cord, 3A
Carrying case (softpack w/shoulder strap)
Shoulder strap (only)
800 ml Disposable collection bottle (48 each)
Filter, 800ml collection bottle, elbow & tubing package
1200 ml Reusable collection bottle (16 each)
1200ml bottle package
Bacteria filter (non-sterile) (12 pack)
6' Patient Tubing
Vacuum gauge test fixture

SET-UP

1. Connect either end of the 4" tubing to the tubing connector then connect the other end to the bacteria filter. Ensure that the clear side of the
bacteria filter is toward elbow and bottle when installing/re-installing. Do not reverse direction of filter.
2. The bacteria filter should then be connected to the 90˚ elbow connection, and the 90˚ connection should be connected to the top of the canister
lid where it says <Vacuum>.
3. The 6' patient tubing should be connected to the canister lid at the outlet labeled <Patient>.
4. Please assure that all connections are secure and without leaks before using.
5. Verify that unit is at desired suction level before beginning patient suction.

HOW TO OPERATE YOUR SUCTION UNIT

Before connecting the unit to the AC adapter or 12V DC cord, make sure that the power switch located on the side of the unit is in the "Off" posi-
tion. Select power source desired.
LED Explanations:
L1- Green– External power supplied to unit from AC power source or DC cord. Illuminated when external power is supplied.
L2 - Yellow– Battery is being charged. Light will go out when battery is fully charged (88 00 20 & 88 00 30 Series only).
L3 - Red– Low battery. Seek another power source and charge battery as soon as possible when light remains on continuous (88 00 20 & 88 00 30
Series only).
AC OPERATION– Plug the small connector of the AC adapter into the DC power input located on the side of the unit. Plug the AC end into a ground-
ed wall-outlet power source.
NOTE– The power supply may become warm to the touch during charging or running of the unit. This is normal.
12V DC OPERATION– (such as a car lighter receptacle). Plug the small connector of the DC power cord into the DC power input on the side of the
unit. Plug the large connector into the 12V DC power receptacle of the automobile.
BATTERY OPERATION– Verify that your unit has an internal rechargeable battery; factory installed on 88 00 20 and 88 00 30 series or provider
installed on 88 00 40 series (installed as an option). To ensure proper operation from internal battery, fully charge the battery for 10-17 hours as
explained in the Battery Charging section. To operate the unit from an internal rechargeable battery, ensure that no external power sources are
plugged into the DC power input on the side of the unit.
Once the power source is selected, simply turn the unit "On" using the power switch located on the side of the unit. The power indicator light (dis-
played on the top of the unit) will indicate which power source is being used by staying continuously lit when external power is connected.
If the unit does not receive an external power source or the battery does not get recharged immediately, the low battery indicator light will
remain on and the performance of the unit will drop off rapidly. Switch to another power source immediately to avoid an interrupted suction
procedure.
100-405-00
A c c e s s o r y & R e p l a c e m e n t I t e m s / S e t - U p / O p e r a t i o n
Part No.
88 44 00
88 44 01
88 44 02
88 44 03
88 44 04
88 44 05
88 45 00
88 45 01
88 46 00
88 46 01
88 47 01
88 47 02
88 48 01
88 48 02
88 49 01
88 49 03
88 50 00

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

88 00 3088 00 40

Table des Matières