Fault Indications; Product Lifetime - arjo Enterprise 9000X Notice D'utilisation

Lits hospitaliers
Masquer les pouces Voir aussi pour Enterprise 9000X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Fault indications

The bed's control software indicates problems in the electrical system by
means of flashing indicators on the Attendant Control Panel (ACP). If you
experience any of the indications below, contact Arjo or an approved
service agent.
Indication
ACP mattress platform height and head
down tilt indicators flashing
ACP mattress platform height and foot
down tilt indicators flashing
ACP backrest indicator flashing
ACP thigh section indicator flashing
Mattress platform height, head down tilt,
backrest and thigh section indicators
flashing

Product lifetime

The lifetime of this equipment is typically ten (10) years. "Lifetime" is defined
as the period during which the product will maintain the specified
performance and safety, provided it has been maintained and operated in
conditions of normal use in accordance with the requirements in these
instructions.
54
Possible cause
Height actuator fault (foot end)
Height actuator fault (head
end)
Backrest actuator fault
Thigh section actuator fault
Control unit fault

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

E9x

Table des Matières