WARNING
Before using patient movement detection, verify that the alarm can be easily heard by
caregivers, e.g. at the nurse's station.
To activate patient movement detection, press either the In Bed button or
Egress button for two seconds.
The threshold display indicator will flash to show the alarm threshold level.
The further to the left the flashing indicator is, the lower the alarm threshold,
and small movements of the patient in bed will be detected.
The further to the right the flashing indicator is, the higher the alarm
threshold, and only large movements, such as the patient exiting the bed,
will be detected.
To increase the patient movement detection threshold:
Press and hold the Egress button; the flashing indicator will move towards
the right. When the required threshold is reached, release the button.
To reduce the patient movement detection threshold:
Press and hold the In Bed button; the flashing indicator will move towards
the left. When the required threshold is reached, release the button.
After a few seconds the indicator will stop flashing and remain lit to show that
patient movement detection is active.
If movement of the patient above the set threshold is detected, an audible
alarm will sound and the threshold indicator will flash.
To cancel the alarm or deactivate patient movement detection:
Press the In Bed button or Egress button once.
The alarm will be silenced and the threshold display indicator will go out to
show that movement detection is deactivated.
Lower
threshold
Higher
threshold
45