Fault indications

Indication
ACP mattress platform height and head
down tilt indicators flashing
ACP mattress platform height and foot
down tilt indicators flashing
ACP backrest indicator flashing
ACP thigh section indicator flashing
Mattress platform height, tilt, backrest and thigh section
indicators flashing.

Product lifetime

The bed's control software indicates problems in the electrical system
by means of flashing indicators on the Attendant Control Panel (ACP).
If you experience any of the indications below, contact Arjo or an
approved service agent.
The lifetime of this equipment is typically ten (10) years. "Lifetime" is
defined as the period during which the product will maintain the
specified performance and safety, provided it has been maintained and
operated in conditions of normal use in accordance with the
requirements in these instructions.
Possible cause
Height actuator fault (foot end)
Height actuator fault (head end)
Backrest actuator fault
Thigh section actuator fault
Control unit fault