Guidance and manufacturer's declaration – electromagnetic immunity
The MC3 motors are intended for use in the electromagnetic environment specified below. The customer or the user of the MC3 motor should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer's declaration – electromagnetic immunity
The MC3 motors are intended for use in the electromagnetic environment specified below. The customer or the user of the MC3 motors should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoret-
ically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the MC3 is used
exceeds the applicable RF compliance level above, the motor MC3 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the MC3 motor.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the MC3
The MC3 motors are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MC3 motors can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the MC3 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power
of the transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
REF 2100195-0002/2014.05/FRA
IEC 60601 test level
±2 kV contact
±4 kV contact
±6 kV contact
±2 kV air
±4 kV air
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±0.5 kV line to line
±1 kV line to line
±0.5 kV line to earth
±1 kV line to earth
±2 kV line to earth
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles <5 % UT
(>95 % dip in UT) for 5 sec
3 A/m
Compliance level
10 V
10 V/m
150 kHz to 80 MHz
0.035
0.11
0.35
1.1
3.5
Compliance level
Electromagnetic environment - guidance
±2 kV contact
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic
±4 kV contact
material, the relative humidity should be at least 30%.
±6 kV contact
±2 kV air
±4 kV air
±8 kV air
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Power frequency magnetic fields should be at levels characteristic of a typical location
30 A/m
in a typical industrial location or hospital environment.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the
MC3 , including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter
Recommende separation distance
where P is the maximum output power rating of the transmitter in watts (W) according to the trans-
mitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
0.035
0.11
0.35
80 MHz to 800 MHz
800 MHz to 2,5 GHz
b
m
1.1
3.5
a
should
800 MHz to 2.5 GHz
0.07
0.22
0.7
2.2
7
3/4