Sources Of Error - Steinberg Systems SBS-LW-200A Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 8
Then press the MODE button. The word PASS flashes. Once the 100,000 figure ap-
pears, put 100g on the scales until the word PASS appears.
Remove the weight and wait till the scale displays 200000. Then place a weight of
200g on the scales until PASS is displayed.
When the word REPEAT appears, repeat the process from the beginning.The process
must be followed by 3 times.
For the scale SBS-LW-300A, you must place first 100 and then 300g on the scales,
which shows 100000 and 300000
6. ATTENTION
1.
Use the scale only as explained above in the technical details.
2.
Use the device on a stable and flat surface. Protect the device from shocks, humi-
dity, radiations and magnetic fields.
3.
If the screen does not show a stable result, make sure the support of the plat-
form is not damaged.
EN
7. RS232 PLUG
Connect the RS232 plug with a mini 16-pin printer. Press the PRINT button to print.
CONNECTION
12
1.
Transfer format: Asynchronous serial communication
2.
Communication Protocoll : Baud-Rate = 9600 Data bit 7 digits, 1 digit bit Starting,
Stopping bit 1 digit
3.
Data Format: Vacant / - sign enter data unit 1digit 7digits 2digits 1 digit All specifi-
cations above are in ASCII.

8. SOURCES OF ERROR

1.
If the balance can not be turned on, check the power source
2.
When the scale does not tare automatically, no stable display is reached. Check
that the platform is well placed.
3.
If the scale displays „HHHHHH" it means there is an overload. Remove the
weight from the platform immediately to avoid a damage of the device. Die Waa-
ge darf nie überlastet werden!
13
EN

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sbs-lw-300a

Table des Matières