General instructions for cleaning and disinfecting the
exterior surfaces of the patient chair M2
The exterior surfaces of the patient chair M2 resist most of the
common surface disinfectants.
However, do not use any
– disinfectants containing concentrated organic or anorganic
acids, since these may cause corrosion damages,
– disinfectants containing chloramides, phenol derivates or
anionic tensides, since these may cause stress cracks in
the plastic material used for the housing of the unit.
Always observe the instructions for use by the
manufacturer of the disinfectants, above all,
concentration specifi cations and instructions.
The chair must be separated from the mains supply
prior to cleaning.
Recommended surface disinfectants
Disinfectant
TERRALIN
(Concentrate)
QUATOHEX
(Concentrate)
Incidin Plus
(Concentrate)
Pursept-A
(Disinfectant spray or
disinfectant cloths)
UK
Ingredients
Benzal conium chloride
Phenoxypropanole
Didecyl dimethyl
ammonium chloride
Benzal conium chloride
Bi-guanidinium acetate
Polymer biguanid
Cleaning agents
Glucoprotamin
Nonionic tensides
Solvents, complexing agents
Ethanol
Glyoxal
QAV
F
(in 100 g)
Manufacturer
20 g
Schülke & Mayr, Norderstedt
35 g
Braun, Melsungen
14 g
10 g
7.5 g
0.5 g
26.0 g
Henkel, Düsseldorf
38.9 g
Merz & Co., Frankfurt/M.
0.1 g
0.05 g