COMPACT PIEZO LED
06.2 Û CLEANING AND STERILIZATION OF THE INSERTS
PREPARATION
- Disconnect the insert from the handpiece
with the torque wrench (see Chapter 05).
EN
MATERIALS NECESSARY
- Enzymatic detergent with pH 6-9
- Water
- Ultrasound tank
- Brush with soft nylon bristles
- Compressed air
- Distilled water
- Single-use sterilization packets
- Steam sterilization
- Syringe
CLEANING AND DISINFECTION METHOD - Inserts
Prepare an enzymatic detergent solution
with pH 6-9, as per the instructions of the
manufacturer;
Place the insert in the ultrasound tank
submerged by the enzymatic detergent
solution at 40°C, for at least 10 minutes;
16
CAUTION: Always disconnect the insert
from the handpiece before proceeding to
its cleaning and sterilization.
CAUTION: Do not use hydrogen
peroxide. If you intend to disinfect the
inserts, do not use hydrogen peroxide, but
only disinfectants with neutral pH (pH7);
always rinse with sterile water to preserve
the disinfection.
CAUTION: Once used, correctly dispose
of the enzymatic detergent solution, do
not recycle.
1
2
10 min
40°C