06 Û CLEANING AND STERILIZATION
06.1 Û CLEANING AND STERILIZATION OF THE HANDPIECE

PREPARATION

- Disconnect the handpiece from the cord of
the dental chair (see Chapter 05);
- If present, disconnect the insert from the
handpiece with the torque wrench (see
Chapter 05);
- Unscrew the front terminal (see Chapter 05).
MATERIALS NECESSARY
- Enzymatic detergent with pH 6-9
- Water
- Container for solution with enzymatic
liquid
- If need be, non-aggressive disinfectant
solution with neutral pH (pH7)
- Clean and soft cloth with low fiber-release
- Brush with soft nylon bristles
- Distilled water
- Compressed air
- Single-use sterilization packets
- Steam sterilization
CLEANING AND DISINFECTION METHOD - Handpiece
06 Û CLEANING AND STERILIZATION
CAUTION: Do not attempt to unscrew
or turn the connector when disconnecting
the handpiece. The connector could get
damaged.
CAUTION: Do not immerge the
handpiece in disinfectant solutions or other
liquids, because it could get damaged.
CAUTION: Do not immerge the
handpiece in an ultrasonic tank.
CAUTION: If you intend to disinfect,
we recommend that you use water-based
disinfectant solutions with a neutral pH
(pH7). Alcohol-based disinfectant solutions
and hydrogen peroxide are contra-
indicated, because they can fade the color
and/or damage the plastic materials. This
also holds true for chemical products such
as acetone and alcohol. Always rinse with
sterile water to preserve the disinfection.
CAUTION: Once used, correctly dispose
of the enzymatic detergent solution, do
not recycle.
Prepare an enzymatic detergent solution
1
with pH 6-9, as per the instructions of the
manufacturer;
EN
13