Pump Indicators The pump front panel has the following indicators:
ON / RESET ALARM
STATIC Mode
Alarm MUTE
The indicator will NOT be illuminated when a Power Fail alarm is muted.
WAIT
HIGH PRESSURE
Hi
mmHg
LOW PRESSURE
mmHg
Lo
Alarm
The green
ON / RESET ALARM
switch is illuminated when the mains power is
POWER
connected and the pump switched on.
The indicator on the
mode has been selected for operation.
Static
The indicator on the
an audible alarm has been silenced.
The
indicator is illuminated when the mattress is
WAIT
being inflated.
The indicator will remain illuminated until the mattress
has been fully inflated. This may take up to 15 minutes.
The
HIGH PRESSURE
the pump detects high pressure within the mattress.
If this condition occurs, the air supply from the pump is
switched off until normal pressure is detected. After 2
seconds of normal pressure being detected the indicator
is switched off and the air supply restarted.
The
LOW PRESSURE
the pump detects low pressure within the mattress.
This may indicate that there is insufficient pressure to
support a patient or that the Transport control is turned
to the
TRANSPORT
connected to the mattress.
The
LOW PRESSURE
once normal pressure is reached.
The pump unit incorporates a sophisticated alarm
detection system that differentiates between patient
movement and genuine alarm conditions.
Whenever an alarm condition is detected the red
triangle starts flashing together with an indicator of the
cause of the alarm. Additionally, an audible warning
will sound, which can be cancelled by pressing the
button (Refer to "Alarm MUTE" on
Alarm MUTE
page 11).
The triangular
Alarm
more of the following indicators:
12
indicator below the
button is illuminated when
STATIC
button is illuminated when
MUTE
indicator is illuminated whenever
indicator is illuminated whenever
position whilst the pump is on and
indicator will be switched off
symbol is displayed with one or
Alarm