Guide and declaration of the manufacturer - electromagnetic immunity
SOPRO device is intended to be used in the electromagnetic environment specifi ed below.
Immunity trial
Conducted RF
3 Vrms
EN 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
EN 61000-4-3
80 MHz to 2,5 GHz
Note 1 : At 80 MHz and 800 MHz, the higher frequency band applies.
Note 2 : These recommendations may not apply in every situation. Electromagnetic wave propagation is modifi ed by the
absorption and refl ection due to the structures, objects and persons.
E N G L I S H
The user should make sure it is used in this environment.
CEI 60601
Compliance
Severity level
level
3V
3V/m
Electromagnetic environment
Guide
Portable and mobile RF communication devices should not
be used at a distance from SOPRO device including the
cables, lower than the recommended separation distance,
calculated with the applicable formulas depending on the
emitter frequency.
Recommended separation distance
d = 1.16√P
d = 1.16√P 80 MHz to 800 MHz
d = 2.33√P 800 MHz to 2.5 GHz
where
is the maximum rated output of the transmitter
in watts (W) by the transmitter manufacturer and d the
recommended separation distance in metres (m).
The fi eld levels emitted by the fi xed RF transmitters,
determined by an electromagnetic measurement of the
sitea, should be lower than the compliance level in each
frequency band b.
Interference may occur in the vicinity of the devices bearing
the following symbol: