SOPRO device is intended to be used in the electromagnetic environment specifi ed below.
Immunity trial
Electrostatic
± 6 kV when in contact
discharges
± 8 kV in the air
EN 61000-4-2
± 2 kV for the feed cables
Far transient bursts
± 1 kV for the input/out-
EN 61000-4-4
put cables
Voltage shocks
Differential mode ± 1 kV
EN 61000-4-5
Common mode ± 2 kV
Dips, brief outages
• <5% Ut - for 10 ms
and power voltage
• 40% Ut - for 100 ms
variation
• 70% Ut - for 500 ms
EN 61000-4-11
• <5% Ut - for 5 s
Magnetic fi eld with the
network frequency
3 A/m
(50/60 Hz)
Note: Ut is the power voltage nominal value applied during the trial.
10
SOPROCARE • User's manual
E N G L I S H
Guide and declaration of the manufacturer - electromagnetic immunity
The user should make sure it is used in this environment.
CEI 60601
Severity level
± 6 kV
± 8 kV
± 2 kV
± 1 kV
± 1 kV
N.A.
<5% Ut 10 ms
<40% Ut 100 ms
<70% Ut 500 ms
<5% Ut 5 s
3 A/m
Compliance
level
The fl oor should be wooden, concrete or tile. If the
fl oor is covered with a synthetic material, the relative
humidity should be at least 30%.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
The main power supply quality should be one of a
traditional commercial or hospital environment.
If the user of SOPROCARE device requires it to continue
to operate during main power supply outages, it is
recommended SOPROCARE device is fed by an inverter
or a battery.
The magnetic fi eld with the network frequency should
be at a characteristic level of a location in a traditional
commercial or hospital environment.
Electromagnetic environment
Guide