In the event of prolonged non use of the device, do not leave it
connected to the motor. Risk of damaging the motor.
V.5 – Connection/disconnection of rotary instrument
r
For all handling of rotary instruments, wear protective gloves. Risk
of cutting and contamination.
x
Only use rotary instruments in good condition. Risk of necrosis.
As the maxi torque given by this MD is very high, this may damage the tool's
driving part. In any case, the compatibility of your instruments has to be
checked. Please follow the manufacturer's instructions and limit the torque
of the unit if necessary. We strongly advise that you use instruments having
a shaft made out of hardened steel.
Check regularly that the push button is set properly by applying a slight
unscrewing torque.
Insertion of rotary instrument
• Apply pressure using your thumb to the centre of the push button and
simultaneously insert the rotary tool up to the stop in the claw of the MD.
• Release the thumb pressure on the push button (Fig 6)
E
For tool replacement: check the good condition of the rotary tool
with light axial traction.
V.6 – Adjustment of the depth stop
It is imperative that the depth stop is adjusted outside of the patient's mouth.
To adjust the depth stop: move the sliding block backward (1) and move the
stop to the required position (2). Loosen the sliding block.
• After each adjustment, check the the stop has been correctly placed using
the ruler.
• Before each use, exercise a slight traction on the stop in order to verify that
it is correctly installed.
• Do not use the depth stop for tapping: the bone may be damaged in the
event of late stop of the driving motor.
The specific depth stops marked AX may only be used for placing Axiom
implants with the Axiom
stop properly in relation to the instrument reference lines.
Sterilisation of medical devices must be done by properly trained
E
and protected staff, in compliance with current regulations. The
sterilisation protocol must be appropriate to the infectious risk.
Wear appropriate protective clothing: risk of infection and injury.
guided surgery. Verify the position of the depth
®
VI.HYGIENE AND MAINTENANCE
®
19